Lead-I ECG for detecting atrial fibrillation in patients with an irregular pulse using single time point testing: a systematic review and economic evaluation

This is a systematic review reporting the estimates of diagnostic accuracy, and cost-effectiveness of lead-I ECG devices. The diagnostic accuracy and clinical impact results presented are derived from an asymptomatic population (used as a proxy for people with signs or symptoms of AF). The summary sensitivity of lead-I ECG devices was 93.9% [95% confidence interval (CI) 86.2% to 97.4%] and summary specificity was 96.5% (95% CI 90.4% to 98.8%). One study reported limited clinical outcome data. Acceptability of lead-I ECG devices was reported in four studies, with generally positive views. The de novo economic model yielded incremental cost-effectiveness ratios (ICERs) per quality-adjusted life-year (QALY) gained. The results of the pairwise analysis show that all lead-I ECG devices generated ICERs per QALY gained below the £20,000–30,000 threshold. KardiaMobile (AliveCor, Mountain View, CA, USA) is the most cost-effective option in a full incremental analysis.